78393402 2024 017 13:00 0 4.01 The old flat protecting window replaced by a new one with coating S1 & S2: AR R < 1% @ 532 nm & 1064 nm & 1550 nm, R < 1% @ VIS, AOI 0 degree 78393402 2023 313 00:00 0 99 A ceiling-mounted cassette air conditioner installed in the laser room, where Clock, ET, Barometer are located as well. 78393402 2023 306 00:00 0 5.02 Pockels cell high voltage driver changed, the crystal remains. 78393402 2023 181 00:00 1 99 Stop sending CRD v1 78393402 2022 180 00:00 1 5.02 A new laser start diode installed; the old replaced 78393402 2021 085 12:00 1 12 New PTU300 (SN: T0210974) with two barameters installed in laser room; old PTU300 (SN: L1110324) for barometric pressure comparison, temperature and humidity 78393402 2021 078 14:00 1 5.02 Lambda/4 re-installed; circular polarization; 78393402 2020 345 00:00 1 12 Old MET3A replaced by a newly calibrated Vaisala PTU300 for Temp, Humidity & Pressure measurements 78393402 2020 062 00:00 0 11 New Aircraft Detection installed (Air!Squitter/JetVision, with ADS-B/MLAT/Flarm) 78393402 2019 070 15:00 1 99 Pulse Distribution Box / Power Supply repaired, cables changed, affects cal constant 78393402 2018 074 00:00 1 99 Pulse Distribution Box / Trigger Level changed; aligned for min jitter; CAL changed by 350 ps 78393402 2015 124 08:00 0 5.02 Laser Beam path on laser table changed; causes CAL constant change only 78393402 2015 023 00:00 0 99 Mount Main dichroic changed to IQOQI 2 (reflects 800 nm, transmits everything else) 78393402 2015 023 00:00 0 99 Mount DetPack New M1/Dichroic (EO86-400): Rmax 532 / Tmax 650-1200 nm 78393402 2015 023 00:00 0 99 Mount Temp stab for C-SPAD housing directly coupled to temp controlled air flow 78393402 2014 289 00:00 0 99 Mount Complete mechanical rebuild of detetction package; but same optical / electronical units 78393402 2014 181 05:00 0 99 RANGE Program updated V7.31: Format#5 for returns; increased for >99999 SoD 78393402 2014 016 11:00 0 99 RANGE Program updated V7.30: FPGA RG FIFO reduced, part of future RGs stored now in PC 78393402 2013 181 00:00 0 99 CPF_WG 2x Iteration changed to 4x Iteration, for horizontal HEO residuals 78393402 2013 171 15:00 0 99 Laser Room at 15:00 UTC; Temperature reduced by 2C to 20C (mount Electronics) 78393402 2013 163 00:00 0 99 Laser Polarization fixed again; Unit disconnected, software still sends commands 78393402 2013 074 00:00 1 8.01 CAL CAL Program extended for different lasers; new setup files; calfile format adapted 78393402 2012 324 00:00 1 5.02 Laser Lambda/4 removed; Lambda/2 RC unit re-installed; linear polarization; restart Glonass Retro Tests 78393402 2012 289 00:00 1 99 Coude Coude mirrors: Cleaned (with Polymer); new airtight path; aligned 78393402 2012 269 00:00 0 4.01 Mount Az Poti included 78393402 2012 254 00:00 0 99 Range NPts_AG: Valid points only if AG is on 78393402 2012 208 00:00 1 99 PDM PDM 2012 installed: Improved NIM; 10kHz ready 78393402 2012 083 00:00 0 5.02 Laser Lambda/4 plate installed; changes linear into circular poarization; starts with LAS08309 78393402 2011 286 00:00 0 99 ET 10 kHz System using Graz RIGA ET again 78393402 2011 262 00:00 1 99 PDM Old (Version red/blue/green front) Pulse Distribution Unit installed: better jitter behaviour then version above; new CAL value 78393402 2011 259 00:00 1 99 PDM Old (Version red/blue front) Pulse Distribution Unit reinstalled: 10 kHz Version - problems with NIM pulse; new CAL value 78393402 2011 257 00:00 0 99 ET 10 kHz System using the San Fernando RIGA ET 78393402 2011 255 00:00 1 99 Start Start Pulse Detector Jan1 installed 78393402 2011 250 00:00 0 5.02 Laser Laser + Start pulse detector new adjustment 78393402 2011 243 00:00 0 99 PDM Pulse Distribution Module: Trigger Level increased from 40 mV to 80 mV 78393402 2011 203 00:00 1 99 TIME HP GPS: Antenna delay was ZERO; new Antenna Delay 153 ns programmed 78393402 2011 180 00:00 0 99 RANGE7.21 Histogram plots: Includes nonlinear noise, higher resolution etc. 78393402 2011 157 00:00 0 99 Filter Blocking filter replaced: OD6 / 93% Transmission / 535 nm CWL, 43 nm BW 78393402 2011 105 00:00 1 99 MCU Error in DE_NANO FPGA removed: Daylight filter now removed again during night 78393402 2011 105 00:00 0 99 Filter Blocked 0.3 nm (35%) filter replaced with unblocked 0.3 nm (48%); CAL quote from 21% to 38%;TFilter: 22 ps 78393402 2011 100 00:00 0 99 Filter Blocking filter (90%, 10 nm) on top of wavelength filter installed 78393402 2011 096 00:00 0 99 MCU DE_NANO FPGA on mount installed; daylight Filter unintentionally not removed during night (but TFilter applied) 78393402 2011 063 00:00 0 4.01 Mount New Dichroic / KV Optical installed; Mount Subroutine: Fraction from 0.5 to 0.2 changed; dramatic improvement 78393402 2011 027 00:00 1 5.02 Laser New Startpulse Detector Version 08-2010 (PDM Version) 78393402 2011 026 00:00 0 99 RANGE 4-Flag-Format 78393402 2011 019 00:00 1 99 Laser Laser box placed on new position; 2 telescopes and old stuff removed; new CAL value 78393402 2011 018 00:00 0 99 RANGE Write to Win_PC: PC104/COM1; Sun Angle; Send_Transp; 78393402 2011 012 00:00 1 99 PDM New Pulse Distribution Unit: 10 kHz Version, with disable / enable bits etc.; new CAL value 78393402 2011 011 00:00 0 5.02 Laser New crystal in Regen Amp; new adjustment;less current in Regen URDM; 5 kV Pockels 78393402 2010 280 00:00 0 5.02 Laser New adjustment, new Pockels Cell, 5 kV, 400 uJ 78393402 2010 264 06:00 0 4.01 FILTER Tubus Temperature Control failure detected; fuse resetted 78393402 2010 253 00:00 0 5.02 Laser Laser Adjustment Problems with new Pockels Cell started; 78393402 2010 245 00:00 0 5.02 Laser New Pockels Cell installed 78393402 2010 200 00:00 0 5.02 Laser Pockels Cell complete readjustment 78393402 2010 181 00:00 0 5.02 Laser Pockels Cell: Delay + Pulse Width adjusted; less trailing pulses 78393402 2010 181 00:00 0 4.01 Mount Elevation: Modules Encoder; Course; Course/Fine replaced with RGO modules 78393402 2010 179 00:00 0 99 ISA FPGA Overlap Check Time: From 60 us increased to 80 us 78393402 2010 159 00:00 0 99 WinClean Version 1.3: NP CRD files routinely stored on DGFI ftp side, parallel to old NPs 78393402 2010 126 00:00 0 99 WinClean Version 1.3: C:Laset2010NPNP126 ...; first CRD NPs (tests only) started 78393402 2010 117 00:00 0 99 CAL CAL opening reduction removed; standard again 78393402 2010 110 00:00 0 6.01.01 SPAD SPAD aligned, using CAL target; CAL opening reduced to 1/3 78393402 2010 105 00:00 0 99 FOV FOV Servo with toothed wheel,for ZERO slackness, no backlash etc. 78393402 2010 036 00:00 0 99 RANGE ET1_Ep + TFilter => ET_TI + TFilter 78393402 2010 036 00:00 0 99 BLITS Filter OUT for BLITS Night Passes 78393402 2010 035 00:00 0 99 RANGE Day/Night Switch (RED LED) instead of LCD; Switch 1/2/4/Speed obsolete (or done via program); 78393402 2009 320 00:00 0 5.02 LASER Adjusted:No Pre-train, a bit more post-tracks 78393402 2009 300 00:00 0 5.02 LASER Oscillator realigned; crazy tracks before (weekend + day 300) 78393402 2009 295 00:00 0 5.02 LASER REGEN + Post readjusted; Indium Seal blocked part of beam; old Regen Pump Diodes reinstalled 78393402 2009 281 00:00 0 99 FILTER OUT for BLS / Night Passes; did not work !!! 78393402 2009 274 00:00 1 99 RG Kabel fuer RG geaendert; RANGE: Del0 von 135 ns auf 160 ns 78393402 2009 273 00:00 0 99 RANGE CO-Filter: IDet_Limit fuer HEOs von 100ps auf 300 ps geaendert 78393402 2009 264 00:00 0 4.01 FILTER OUT for Night/HEO/AUTO: 15 ps added for missing filter glas 78393402 2009 246 00:00 0 99 RANGE Mount_Loop: 30 ms/100 ms Update Rates for HEO/LEO Satellites 78393402 2009 223 00:00 0 4.01 ISIT RCA ISIT (US system) defect, JAI ISIT installed (European System); no Seeing values derived 78393402 2009 195 00:00 0 5.02 LASER Pockels Cell readjusted; Leading edge of rets back to normal; energy not stable; 1-s CHA intervals ? 78393402 2009 194 00:00 0 5.02 LASER REGEN: Pump Diodes replaced; Pockels Cell adjusted; increased jitter @ strong returns 78393402 2009 188 06:00 0 99 FPGA Times Before: 165.66-29.95=135.71 ns; After: 145.23-25.33=119.90 ns 78393402 2009 188 00:00 0 99 ISA FPGA RGG: Min/Max Delay corrected; before: up to 20 ns drift / 100 ns interpolation (analog chip) 78393402 2009 183 00:00 0 99 RANGE Mount_Loop: From 30 ms to 100 ms update intervals changed 78393402 2009 182 00:00 0 99 Station Thunderstorm / Flash: Serial Interface to Dome / Dome Frequency Converter damaged 78393402 2009 105 00:00 0 99 Station Power failure for some hours 78393402 2009 061 00:00 0 99 PC CPU Card defect; exchanged .... 78393402 2009 043 00:00 0 99 OCU OCU and MCU operational 78393402 2009 015 00:00 1 99 Format Epochs: From F14.8 to F18.12 (10 ns to 1 ps; only 10-100 ps acc); WinClean, Range changed 78393402 2009 014 00:00 0 99 SatDef C:LasetDefSatDef replaces all other sats/sat.cpf etc. files; all progs changed 78393402 2009 014 00:00 0 99 Dome Observer Cabin: New Table / PCs disconnected / new connections etc. 78393402 2008 342 00:00 0 99 Dome Station up; new dome / RS232 control implemented in Range 78393402 2008 311 00:00 0 99 Dome Station down; Old Dome Removed; all cables OFF etc 78393402 2008 198 00:00 0 99 Transmit Testing New Transmit Telescope (20 cm lens); no data sent 78393402 2008 197 00:00 0 99 CPF Predicts now with New PC/XP, also the ftp transfer / Zipped 78393402 2008 178 00:00 0 99 KG PC New PC/XP; Softcal etc. LF9045 78393402 2008 161 00:00 0 99 MET PC New PC / XP 78393402 2008 065 00:00 1 99 WinClean AJI: LE /20 mm/ 5 mm RMS started 78393402 2008 063 00:00 0 5.02 Start Start Pulse Detector adjusted: From 2.5 V to 3.5 V 78393402 2008 059 00:00 1 5.02 CAL Start Pulse Detector: BNC/SM/BNC changed to SMA only; 200 ps jump in CAL value 78393402 2008 047 00:00 0 99 CLEAN Error Writes to Statistics corrected: Pts/NP; Quote were wrong ... 78393402 2008 043 00:00 0 6.01.01 SPAD Realigned; Laser realigned; now much better results 78393402 2008 036 00:00 1 99 WinClean LA1/LA2: Leading Edge /20 mm/ started (5mm RMS); AJI: LE /7 mm RMS / started 78393402 2008 023 00:00 0 99 CLEAN New WinClean Program operational; mainly doing manual, but also few AutoCleans 78393402 2007 263 00:00 0 12.03.01 MET External analog MET sensors transferred into Laser Room 78393402 2007 262 00:00 0 6.01.01 ET ET_Sync; maybe power failure; thunderstorm 78393402 2007 261 00:00 0 99 GPS HP GPS Antenna: Back to (roughly) previous location; restart 78393402 2007 235 06:00 0 99 Mount Power Supply / Sun Shutter etc. failure / worked again ... 78393402 2007 234 00:00 0 99 Mount Power Supply Failure; repaired; fuse 78393402 2007 234 00:00 0 99 Mount Several Failures due to frequent OFF/ON (observatory) 78393402 2007 211 00:00 0 10.01 CLEAN Results: Noise points showed up to now an offset of +1.3 m (SemiTrain Distance ...); corrected 78393402 2007 205 00:00 0 99 GPS HP GPS Antenna: New Location: Roof Middle; + 403 cm higher, same cable length, switched OFF/ON 78393402 2007 177 00:00 0 99 LAS PC All relevant Programs with COM_Select (NN): TWO Writes to CAS-81 instead of ONE; works again 78393402 2007 170 00:00 0 99 LAS PC CF Card replaced by standard disk again 78393402 2007 169 00:00 0 99 LAS PC CF Card 2 GB used instead of Hard Disk; all systems on CF card 78393402 2006 258 00:00 0 99 Mount Cabl RG58/SMA Stop1 from SPAD to Base Plate installed 78393402 2006 250 00:00 1 99 Mount Cabl New RG58/BNC cable from SPAD to Contraves base plate (flexible core) instead of RG223 (rigid core) 78393402 2006 249 00:00 0 99 Mount Cabl jumps; tests with fiber; due to cable/AZ/crosstalk? cable ties changed; better, but not solved 78393402 2006 241 09:00 0 99 Mount Cabl 9:00 Same jumps again(60 ps): Contact spray on Az Base SMA Connectors, Stop4; 78393402 2006 241 00:00 0 99 Int PCB Temp Control reconnected; +- few ps again 78393402 2006 240 12:10 0 99 Mount Cabl 12:10 SMA Cable for Stop Pulse: Changed from Stop1 to Stop3 SMA Cable 78393402 2006 237 16:00 0 99 Mount Cabl 16:00 4 SMA cables RG223 at yoke from outside touched, big cal variations; after that okay ... 78393402 2006 236 14:30 0 6.01.01 C-SPAD 14:30 => Internal PCB Temp Control (Heating) disconnected. T decreases => Cal decreases 78393402 2006 233 13:00 0 6.01.01 C-SPAD From 233 - 234 / 13:00 UTC: CAL var +-40 ps, Temp Sensor SPAD Housing "obigfoin" 78393402 2006 212 00:00 1 5.02 kHz Laser Start Pulse aligned; 3 days before bigger variations; temp sensor was not on C-SPAD anymore 78393402 2006 191 00:00 0 99 CPF Pred CPF predictions; fully automatic preds since day 198 78393402 2006 053 00:00 0 5.02 kHz Laser Laser readjusted, REGEN Pump increased by 0.2 A; Osc. decreased by 100 mA 78393402 2006 045 00:00 0 99 kHz Laser Start Puls Detector Position readjusted; from +2V to +3V pulses 78393402 2006 016 00:00 0 99 LAS PC Switch to Industry PC/3.2 GHz, with IEC-Bus/LPM16/DIO96 Cards; No ISA/Altera 78393402 2005 314 00:00 0 99 MET PC HD Failure during rearrangements; on evening everything restarted again 78393402 2005 312 00:00 0 5.02 kHz Laser Oscillator Adjust.; complete adjust; 1050 to 900 mA Osc; 375 Units 78393402 2005 311 00:00 0 99 Laser Room 2 Racks removed; all devices in Rack#3;(Rack4: Mount); IEC-Bus: Needs 5328 78393402 2005 297 00:00 0 5.02 kHz Laser Osci Dioden same; nachjust no change; Start Pulse: from 2.7 => 3.0 V 78393402 2005 290 06:00 0 5.02 kHz Laser Oscillator Pump Diodes: +50 mA; RMS slightly degraded then 78393402 2004 212 00:00 0 99 CALK After Laser Readj.(208): 5 Round trips instead of 6-7, RG was slightly too late; corrected now 78393402 2004 210 00:00 1 12.01.01 MET3A Recalibrated; installed again; Before: 0.290 hPa diff; now: 0.188 hPa diff 78393402 2004 208 00:00 0 5.02 kHz Laser Schmidt: Complete new alignment; new crystal postamp; C-SPAD aligned; RMS CAL from 14 now 20 ps 78393402 2004 189 00:00 0 12.01.02 MET3A Removed for Recalibration; No MET-PC Records; Using Paroscientific + external sensors 78393402 2004 093 00:00 0 6.01.01 SPAD PS New Spad Power Supply installed; PS3-10; uncompensated Out now via RG224 Stop #2 78393402 2004 058 00:00 0 6.01.01 TWC TWC new adjustment; from 110 ps to <10 ps; Power Supply Poti Problem 78393402 2004 028 00:00 0 5.02 kHz Laser Mr. Frei (HQ) fixed the crystal of the RegenAMP and the 2.AMP; 78393402 2004 028 00:00 0 5.02 New Alignment (RM1, RM2 etc.; Start Pulse Sensor); from 840 AU up to 940 AU; 78393402 2004 021 00:00 1 5.02 kHz Laser Spad Unit 0406 / Test Ch. Timewalk new adjusted PS: -6V to -6,450V 78393402 2004 021 00:00 0 5.02 kHz Laser SESAM: from 1000mA to 700mA no change in energy, timewalk and cal-time 78393402 2004 021 00:00 0 5.02 kHz Laser New Alignment (RM1, RM2 etc.); from 810 AU up to 910 AU; 78393402 2004 016 00:00 0 5.02 kHz Laser SESAM: from 700mA to back 1000mA ?SESAM, pulse length / time walk?? 78393402 2004 015 00:00 0 5.02 kHz Laser pulse distribution box reinstalled; Triggervalues: Start=400mV, Stop=150mV 78393402 2004 013 00:00 0 5.02 kHz Laser SESAM: from 1000mA to 700mA no change in energy and cal-time 78393402 2004 013 00:00 0 5.02 kHz Laser max energy on start pulse detector (lower timewalk for energy changes, warmup) 78393402 2004 013 00:00 0 5.02 kHz Laser pulse distribution box removed (defect) 78393402 2004 007 00:00 0 12.03.02 kHz Laser New Alignment (RM1, RM2 etc.); from 760 AU up to 890 AU; 78393402 2003 345 00:00 0 9.02.01 GPS HP GPS PS fan 5V Supply only (Oven could not stabilize) 78393402 2003 331 00:00 0 6.01.01 TWC SPAD Time Walk Compensation: Adjusted for 10 ps kHz Laser 78393402 2003 331 00:00 0 9.02.01 GPS HP GPS new power supply installed (incl. 12V fan) 78393402 2003 324 00:00 0 5.02 kHz Laser New Alignment (RM1, RM2 etc.); from 0.25 mJ up to 0.40 mJ again; 78393402 2003 195 00:00 0 99 RGG RGG2, with FIFOs (512); Range 5.40; runs up to 1 kHz (NO Plot_Resids, but WITH MOUNT etc.) 78393402 2003 170 00:00 0 99 Computer SPECTRA 2.4 GHz ISA Pentium-4 Board; Running Range 5.31; Free running, MJS RGG 78393402 2003 015 06:00 0 12.01.01 PARO ParoScientific Digiquartz: New Calibration, running parallel, visual comparison 78393402 2002 344 00:00 0 12.01.01 MET3A RANGE/CAL/MET: Using MET3A ONLY + PC_Met; ParoScientific DigiQuartz => Calibration 78393402 2002 331 00:00 0 99 RANGE RANGE 5.31: Free running, MJS-RGG; 2xSR620 read/Store; CAL: 10-Hz-Version 78393402 2002 330 00:00 0 99 DOME New floor at Dome; all cables dis-/re-connected 78393402 2002 330 00:00 0 6.01.01 C-SPAD TWC new adjustment; -98 ps @ 1000 Photons before(!!!); now +3 ps; see SPAD Unit 406 78393402 2002 322 00:00 0 99 RANGX AutoSwitch for CAL (DG535/OLD); Laser Control Bit at F3 instead of Energy Control 78393402 2002 316 00:00 0 99 RANGX All Passes now with Free Running Range/RGG/10Hz ALL (5.30); CAL with DG535/OLD 78393402 2002 255 00:00 0 99 PENTIUM Pentium 200 MHz replaces PC486/66MHz 78393402 2002 183 00:00 0 99 CABLES 4 new cables RG223 (SMA1-SMA4, for STOP1-STOP4); new CAL Constant 78393402 2002 182 00:00 0 99 CABLES Test Channel / Stop 1: New Cable 78393402 2002 170 00:00 0 99 RANGE SPAD Gate Safety: From 120 nm to 51 ns (as previously) 78393402 2002 163 00:00 0 6.01.01 Filter New Filter (Andover 2002, 0.3 nm, 30%) installed 78393402 2002 162 00:00 0 5.01 Laser Nd:YAG, 532 nm, 35 ps, SemiTrain, 30 mJ 78393402 2002 162 00:00 1 6.01.01 Detector C-SPAD, 200 um, cooled to -60 degrees 78393402 2002 162 00:00 0 6.01.01 Counter ET; Clock Module, 1 Start, 1 Stop Module; 78393402 2002 162 00:00 0 99 Computer PC-486, 1xHPIB, Analog/Dig I/O; 78393402 2002 162 00:00 0 99 Range Gate Via HP5359 + DG535 78393402 2002 162 00:00 0 99 Software V 5.3, Tandems, OMO/GEO Mount system, Lagrange 78393402 2002 162 00:00 0 9.01.01 Timing 1Hz / 10 MHz via HP GPS 78393402 2002 162 00:00 0 12.02.01 MET ParoScientific #1; Home made temp/Hum sensors 78393402 2002 137 00:00 0 6.01.01 ET Module 3 disconnected 78393402 2001 094 00:00 0 5.01 LASER Oscillator + Amplifiers + Power GROUNDED 78393402 2001 031 00:00 0 6.01.01 ET From CH1 to Ch6 78393402 2001 030 00:00 0 9.01.01 HP5370A Fan broken; changed from 110 V AC to 220 V AC 78393402 2001 029 06:00 0 9.02.01 GPS New Power Supply installed for HP GPS 78393402 2001 025 00:00 0 6.01.01 ET Back to Ch1; new PCB ECL to TTL 78393402 2001 019 00:00 0 6.01.01 ET ET: From Ch1 to Ch3 78393402 2001 005 00:00 0 99 POWER Electrical Power failure; 3 h no power; ET synced 78393402 2001 005 00:00 0 9.02.01 GPS HP GPS Power Failure; external Supply 78393402 2001 001 00:00 0 6.01.01 ET New Synced 78393402 2000 214 00:00 0 6.01.01 ET ET ONLY; no counters; HP remains for Checks, sync; 78393402 2000 018 00:00 0 99 COUNTER Jitter in Counter Diffs; Change of 10 MHz Distr.; (KF SLR Timing) 78393402 2000 010 00:00 1 9.01.01 TIME UTC(TUG) disconnected by D. Kirchner; HP GPS now used 78393402 1999 328 00:00 0 5.01 START New PCB Start Pulse Detekt. + PrePulseDetekt. 78393402 1999 307 00:00 0 5.01 START New Position; new Cabeling of Start Pulse Detektor SFH229 78393402 1999 293 00:00 0 6.01.01 SPAD Heating Unit MOunt Detection Package 78393402 1999 291 00:00 0 99 CABLES New Cabeling f New Comparator Unit 78393402 1999 288 00:00 0 99 COMPAR New Comparator Unit; Delay DG535=100 ns; 78393402 1999 278 00:00 0 99 COUNTER COMP1/COMP2 disconnected from computer/OUT/IN/RECON. 78393402 1999 270 00:00 0 5.01 START New Start Pulse Detector SFH229 78393402 1999 253 00:00 0 99 Air Cond Sensor moved to Air Exit 78393402 1999 062 00:00 0 5.01 DYE Complete new Dye 78393402 1999 061 00:00 0 5.01 DYE CELL Flowing Dye Cell; Semitrain Distance 5752 ps 78393402 1999 054 00:00 0 5.01 SEMITRAIN Distance now: 8757 ps (before: 8759 ps) 78393402 1998 349 00:00 0 99 COUNTER CAL/RANGE: HP5370 as reference again 78393402 1997 364 00:00 0 99 PREPY RG: from 1 to 6 us suggested start value 78393402 1997 343 00:00 0 99 CABLES OUT1-Test: BNC4 ==> BNC8 78393402 1997 343 00:00 0 99 Air Cond RH_Int: From 2% to 22% (Poti ZERO) 78393402 1997 168 00:00 0 6.01.01 SPAD TWC new adjustment 78393402 1997 132 00:00 0 6.01.01 SPAD 29.5 ==> 29.8 V 78393402 1997 126 06:00 0 99 COUNTER SR620#3 added again 78393402 1997 114 00:00 0 99 COUNTER SR620#3 removed; repair 50 Ohm Stop 78393402 1997 112 00:00 0 99 FORMAT CLEAN: Format 1 implemented 78393402 1997 073 00:00 0 99 PC PentiumPro 200 MHz at Office, for Cleans etc. 78393402 1997 059 00:00 0 99 FORMAT Range/Clean: Format 1: Color/NTrack/Epoch/4xCounter 78393402 1997 057 00:00 0 9.01.01 TIME Sync Divider: Changes in 1 kHz Epoch Timing 78393402 1997 042 00:00 0 6.01.01 SPAD Daylight filter added 78393402 1997 041 00:00 0 6.01.01 SPAD dismounted, dried, mounted, adjusted; no daylight filter 78393402 1997 034 00:00 0 99 CAL/RANGE HP/SR620A results exchanged; SR620A now used as reference 78393402 1997 030 00:00 0 99 TIME Back to UTC(TUG) again ... 78393402 1997 023 00:00 0 99 TIME HP GPS DEMO used 78393402 1996 355 00:00 0 6.01.01 SPAD STOP SPAD STOP Somp Box Trigger: 250 ==> 100 mV (25 ==> 20 ps) 78393402 1996 334 00:00 0 6.01.01 Trigger Pre-Trigger-Unit adjusted, CAL now okay again 78393402 1996 296 00:00 0 99 Counters SR620C removed 78393402 1996 271 00:00 0 10.01 NP Min 5 Pts/NP minimum, otherwise NP not delivered 78393402 1996 271 00:00 0 6.01.01 TWC New, excellent adjustment 78393402 1996 269 00:00 1 99 Counters No longer f-daisy-chaining of counters; separate 10-MHz Supplied 78393402 1996 260 00:00 0 6.01.01 TWC Jitter after 6 days interruption 78393402 1996 254 00:00 1 6.01 SR620C Added; all 4 counters in use; 78393402 1996 253 00:00 0 99 PC IEC Bus cables changed to PC-Cables 78393402 1996 215 00:00 0 6.01 SPAD HP3631 Power Supply for SPAD; Remot Control via SPAD ON 78393402 1996 198 00:00 0 6.01.01 Trigger New Trigger Devices / MultiOutputs for Start/Stop 78393402 1996 170 00:00 0 99 Air Cond Air Temperature Control for hp only; SR620A/B: orig. fan setting 78393402 1996 144 00:00 0 12.01.02 ParoScientific Sensor: +21 cm higher; software corrected 78393402 1996 143 00:00 0 99 Temp Controlled Air for 3 Counters Cooling 78393402 1996 133 00:00 0 99 New Cabling of Mount/SPAD/Controls 78393402 1996 107 06:00 0 8.01 Target03 used, beginning 06:00 UTC 78393402 1996 060 00:00 0 4.01 Daylight filter installed again in Standard Channel 78393402 1996 059 00:00 0 6.01.01 SPAD adjusted again 78393402 1996 058 00:00 0 6.01.01 SPAD+TWC: From Test into Standard Channel 78393402 1996 054 00:00 0 8.01 Softcal, using Calfile 78393402 1996 046 00:00 0 8.01 CAL: Actual Met Vals used; CalFile Option available 78393402 1996 030 00:00 1 6.01.01 Clean routinely with 3 Counters 78393402 1996 025 00:00 1 6.01.01 CAL/RANGE: 3 Counters used, Results stored; not yet in CLEAN 78393402 1995 269 00:00 0 99 Sat_Char*3: New Pass Names 78393402 1995 265 00:00 1 12.01.01 ParoScientific Pressure Sensor installed 78393402 1995 261 00:00 0 6.01.01 SPAD VOltage set via PC Program 78393402 1995 238 00:00 0 99 USNO correction now also applied in PC 78393402 1995 234 00:00 0 99 First pass with PC 78393402 1995 016 00:00 0 6.01.01 HP5370A: 0.25/0.21 V Trigger Threshold (Before: 0.25/0.17V)