Date: | 2008-10-16 09:16:00 | |
Sender: | W. Gurtner <Werner Gurtner <werner.gurtner@aiub.unibe.ch>> | |
Subject: | [SLR-Mail] No. 1723: Zimmerwald Range Bias Problem | |
Author: | W. Gurtner | |
Content: | ******************************************************************************** SLR Electronic Mail 2008-10-16 09:16:00 UTC Message No. 1723 ******************************************************************************** Author: W. Gurtner Subject: Zimmerwald Range Bias Problem According to the latest range bias reports Zimmerwald developed a range bias problem of about 7 cm on Friday October 10, 2008. We observed at the same time a slight increase in the single shot RMS, the terrestrial target calibration value however did not change. There was some maintenance work done on this day, obviously correlating with the apparent bias appearing in the reports from this day on. We have stopped our SLR activities till after the laser workshop. Werner Gurtner From: Werner Gurtner ******************************************************************************** |